<strong>Surgery</strong> Manual<br />
ROOT-LINE Insertion<br />
You have three options for the final<br />
insertion of the implant:<br />
- Power-assisted insertion (A)<br />
- Manual insertion with the torque<br />
wrench and driver (B)<br />
- Manual insertion with the cardanic<br />
driver (C)<br />
During insertion, make sure to use the<br />
external mark for the groove position in<br />
the implant.<br />
84<br />
A<br />
A. Power-assisted insertion with the<br />
driver for screw implants, ISO shaft<br />
(max. 15 rpm).<br />
If the bone quality is D1 or D2 at the<br />
implant insertion site, the implant must<br />
be externally cooled (saline solution,<br />
5°C/41°F) during the insertion.<br />
B C<br />
B. Manual insertion with the locked<br />
torque wrench and driver (short/long)<br />
for screw implants.<br />
C. Manual insertion with the cardanic<br />
driver.