<strong>Surgery</strong> Manual<br />
CYLINDER-LINE Insertion<br />
Implant Insertion<br />
The implant is removed from the primary<br />
package (blister) by the sterile handle.<br />
Contamination from non-sterile parts<br />
must be avoided.<br />
WARNING!<br />
Prior to insertion in the implant<br />
bed, the implant must be carefully<br />
aligned to the correct groove position<br />
with the aid of the external<br />
marks on the implant body.<br />
Changes cannot be made later!<br />
Pressing the handle into the<br />
implant bed<br />
132<br />
Lifting off the handle<br />
Taking the implant by the handle, insert<br />
it into the implant bed. Make sure to<br />
follow the orientation of the implant bed<br />
axis. Then press the implant in until it<br />
has sufficient grip to permit withdrawal<br />
of the handle. Take care that the tilt-off<br />
placement head is not bent too soon.<br />
Optional Accessory<br />
<strong>CAMLOG</strong> ® PickUp Instrument<br />
If the intraoral space is insufficient to<br />
allow insertion of the implant with the<br />
handle, the implant can be separated<br />
from the handle with the PickUp instrument.<br />
Slide the PickUp instrument<br />
between the implant and the handle<br />
onto the tilt-off placement head and lift<br />
off the handle. The implant is then<br />
inserted into the bone and the PickUp<br />
instrument is withdrawn.