<strong>Surgery</strong> Manual<br />

CYLINDER-LINE Insertion<br />

Implant Insertion<br />

The implant is removed from the primary<br />

package (blister) by the sterile handle.<br />

Contamination from non-sterile parts<br />

must be avoided.<br />

WARNING!<br />

Prior to insertion in the implant<br />

bed, the implant must be carefully<br />

aligned to the correct groove position<br />

with the aid of the external<br />

marks on the implant body.<br />

Changes cannot be made later!<br />

Pressing the handle into the<br />

implant bed<br />

132<br />

Lifting off the handle<br />

Taking the implant by the handle, insert<br />

it into the implant bed. Make sure to<br />

follow the orientation of the implant bed<br />

axis. Then press the implant in until it<br />

has sufficient grip to permit withdrawal<br />

of the handle. Take care that the tilt-off<br />

placement head is not bent too soon.<br />

Optional Accessory<br />

<strong>CAMLOG</strong> ® PickUp Instrument<br />

If the intraoral space is insufficient to<br />

allow insertion of the implant with the<br />

handle, the implant can be separated<br />

from the handle with the PickUp instrument.<br />

Slide the PickUp instrument<br />

between the implant and the handle<br />

onto the tilt-off placement head and lift<br />

off the handle. The implant is then<br />

inserted into the bone and the PickUp<br />

instrument is withdrawn.