11.01.2013 Views

1 Surgery CAMLOG Compendium

1 Surgery CAMLOG Compendium

1 Surgery CAMLOG Compendium

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>Surgery</strong> Manual<br />

ROOT-LINE Insertion<br />

Form Drilling<br />

Diameter and Lengths<br />

Diameter- and length-calibrated form<br />

drills are available for each implant size.<br />

The internally irrigated form drills are<br />

color-coded and laser-marked.<br />

The form drills included in the surgery<br />

sets are supplied with a color-coded<br />

removable depth stop (height 2 mm).<br />

Depending on the defined drilling depth<br />

(implant length), the hole diameter is<br />

expanded progressively with the series<br />

of form drills until the planned implant<br />

diameter is reached.<br />

Recommended drilling speeds:<br />

Ø 3.8 mm 500 rpm<br />

Ø 4.3 mm 400 rpm<br />

Ø 5.0 mm 350 rpm<br />

Ø 6.0 mm 300 rpm<br />

The protocol-compliant insertion depth<br />

for ROOT-LINE Implants with depth stop<br />

in place is the implant length (9 / 11 /<br />

13 mm) minus 0.4 mm with the circular<br />

stop at the crestal bone level so that the<br />

implant shoulder extends 0.4 mm above<br />

the bone level.<br />

If a greater insertion depth is desirable,<br />

the depth stop can be removed and the<br />

preparation will then be guided by the<br />

depth mark.<br />

74<br />

Ø 3.8 mm Ø 4.3 mm Ø 5.0 mm Ø 6.0 mm<br />

ROOT-LINE form drills<br />

Length<br />

9 mm 11 mm 13 mm 16 mm

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!