1 Surgery CAMLOG Compendium
1 Surgery CAMLOG Compendium
1 Surgery CAMLOG Compendium
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>Surgery</strong> Manual<br />
ROOT-LINE Insertion<br />
ROOT-LINE Tap<br />
Introduction<br />
All ROOT-LINE Implants come with a selftapping<br />
thread with sand-blasted, etched<br />
surface (Promote ® ). Use of a tap is recommended<br />
with bone qualities D1 or D2.<br />
The taps are internally irrigated. The maximum<br />
drilling speed should not exceed 15<br />
rpm during power-assisted tapping. We<br />
recommend manual tapping.<br />
ROOT-LINE, SCREW-CYLINDER-LINE tap<br />
Ø 3.8 mm Ø 4.3 mm<br />
78<br />
max. 15 rpm<br />
Ø 5.0 mm Ø 6.0 mm