Manual tapping is performed with the<br />

adapter with ISO shaft and the locked<br />

torque wrench. Make sure that the<br />

implant bed axis orientation is maintained<br />

during insertion and removal of<br />

the tap. The limit for insertion of the tap<br />

is the upper edge of the cutting blade.<br />

See also the instructions for the torque<br />

wrench on page 151.<br />

Caution! The SCREW-CYLINDER-LINE<br />

taps must not be used with SCREW-<br />

LINE Implants.<br />

Note<br />

Since no tapping is required in the<br />

maxilla for bone qualities D3 or<br />

D4, we show a situation in the<br />

mandible for demonstration purposes.<br />

Locked torque wrench<br />

Adapter, ISO shaft<br />

<strong>Surgery</strong> Manual<br />

SCREW-CYLINDER-LINE Insertion<br />

105