1 Surgery CAMLOG Compendium
1 Surgery CAMLOG Compendium
1 Surgery CAMLOG Compendium
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Accessories<br />
Adapter for Screw Implants, long,<br />
for narrow Gaps<br />
A special reduced-diameter adapter for<br />
screw implants, long, is available for<br />
implant diameter 3.8 mm to assist manual<br />
insertion into small gaps. The set-up<br />
must be performed under sterile conditions.<br />
After removal from the primary<br />
package (blister) the color-coded holding<br />
sleeve with correct diameter is inserted<br />
over the endosseous part of the implant<br />
to guide the adapter for screw implants,<br />
long, into the screw implant. The holding<br />
sleeve is squeezed together like a collet<br />
at the level of the implant shoulder and<br />
the handle is removed. Next, the insertion<br />
post is unscrewed with the screwdriver<br />
and removed.<br />
Setting up the holding<br />
sleeve<br />
Adapter for screw implants,<br />
long, Ø 3.8 mm<br />
Withdrawing<br />
the handle<br />
Unscrewing the<br />
retention screw<br />
<strong>Surgery</strong> Manual<br />
ROOT-LINE Insertion<br />
Holding sleeve for implants to guide the<br />
adapter for screw implants, long, into the<br />
screw implant, Ø 3.8 mm<br />
Removing the<br />
insertion post<br />
87