Accessories<br />

Adapter for Screw Implants, long,<br />

for narrow Gaps<br />

A special reduced-diameter adapter for<br />

screw implants, long, is available for<br />

implant diameters 3.3 and 3.8 mm to<br />

assist manual insertion into small gaps.<br />

The setup must be performed under sterile<br />

conditions. After removal from the<br />

primary package (blister) the color-coded<br />

holding sleeve with correct diameter is<br />

inserted over the endosseous part of the<br />

implant to guide the adapter for screw<br />

implants, long, into the screw implant.<br />

The holding sleeve is squeezed together<br />

like a collet at the level of the implant<br />

shoulder and the handle is removed.<br />

Next, the insertion post is unscrewed<br />

with the screwdriver and removed.<br />

Adapter for screw implants,<br />

long, Ø 3.3 / 3.8 mm<br />

Setting up the holding sleeve Withdrawing the handle Unscrewing the retention<br />

screw<br />

<strong>Surgery</strong> Manual<br />

SCREW-LINE Insertion<br />

Holding sleeve for implants to guide the<br />

adapter for screw implants, long, into the screw<br />

implant, Ø 3.3 mm and Ø 3.8 mm<br />

Removing the insertion<br />

post<br />

61