1. Initial Notes<br />

NOTES<br />

• The current user manual is only valid for<br />

VDW.GOLD ® devices upgraded with Software<br />

version R 1.0. With this Software<br />

Upgrade, the VDW.GOLD ® endo motor can<br />

be used both in rotary as well as in reciprocating<br />

mode. If you have upgraded your<br />

VDW.GOLD ® with Software version R 1.0,<br />

please replace your current VDW.GOLD ®<br />

user manual with these instructions.<br />

• For detailed instructions on how to upgrade<br />

your VDW.GOLD ® with Software version<br />

R 1.0, please refer to the separate<br />

Upgrade Instructions in your RECIPROC ®<br />

Upgrade Kit.<br />

• For detailed instructions on how to use<br />

RECIPROC ® instruments, please refer to<br />

the RECIPROC ® DIRECTIONS FOR USE enclosed<br />

in your RECIPROC ® Upgrade Kit or<br />

consult www.reciproc.<strong>com</strong>.<br />

• If you have already used your VDW.GOLD ®<br />

device, the initial steps for preparing and<br />

installing the device described in these<br />

instructions might not be relevant for you.<br />

As you have added a new function to your<br />

endo motor, we re<strong>com</strong>mend, however, to<br />

read these directions for use <strong>com</strong>pletely<br />

and thoroughly.<br />

2. Indications for Use<br />

FOR DENTAL USE ONLY!<br />

VDW.GOLD ® upgraded with Software R 1.0 is a<br />

medical device in accordance with the Medical<br />

Device Directive 93/42/EEC, amended with the<br />

Directive 2007/47/EC. The endo motor is exclusively<br />

designed for use by dentists with <strong>dental</strong><br />

root canal instruments in continuous rotation and<br />

in reciprocating movement with integrated apex<br />

locator (endometric length determination).<br />

This device may only be used by qualified <strong>dental</strong><br />

professionals in hospital settings, clinics or <strong>dental</strong><br />

practices.<br />

41<br />

3. Contraindications<br />

Do not use the device for implants or other nonendodontic<br />

<strong>dental</strong> procedures.<br />

4. Warnings<br />

This chapter contains a description of severe<br />

adverse reactions and potential safety hazards for<br />

the product or the user/patient. Please read the<br />

following warnings before use.<br />

WARNINGS<br />

Upgrade<br />

• The RECIPROC ® Upgrade to Software R<br />

1.0 will reset the VDW.GOLD ® , meaning<br />

the device will return to its factory default<br />

parameters. This reset will also include deletion<br />

of all individual Dr’s Choice settings.<br />

Users<br />

• The device is designed for use in <strong>dental</strong><br />

care and may only be operated by trained,<br />

qualified professionals, e.g. dentists.<br />

Ambient Conditions<br />

• The device should not be placed in damp<br />

areas or anywhere where the device may<br />

<strong>com</strong>e into contact with any type of liquids.<br />

• Do not expose the device to any direct or<br />

indirect sources of heat. The device must<br />

be operated and stored in a safe environment.<br />

• Do not use the device in the presence<br />

of free oxygen, anaesthetic substances or<br />

flammable products. The device must be<br />

operated and stored in a safe environment.<br />

• The device requires special safety precautions<br />

with regard to electromagnetic<br />

<strong>com</strong>patibility (EMC), and the EMC information<br />

contained in this manual must be<br />

strictly adhered to during installation and<br />

operation. It is especially important not<br />

to use the device near fluorescent lamps,<br />

radio transmitters or remote controls,<br />

portable or mobile radio frequency <strong>com</strong>munications<br />

equipment.<br />

• In order to avoid possible danger from<br />

electromagnetic interference, do not operate<br />

any other electromedical or other electrical<br />

en<br />