68<br />

Problem Possible Cause Solution<br />

en<br />

APEX STATUS LED<br />

does not switch to:<br />

red (stays off):<br />

green:<br />

Faulty connection between the<br />

lip clip cable and the unit.<br />

Defective lip clip cable.<br />

Lip clip and contra-angle file<br />

or lip clip and separate file clip<br />

connected too fast or incorrectly.<br />

• Contra-angle not functioning correctly.<br />

• Check whether the lip clip cable is<br />

inserted incorrectly or whether it is<br />

twisted or defective.<br />

• Clean the lip clip.<br />

• Check whether the measuring file<br />

has been correctly inserted into the<br />

contra-angle.<br />

• Check whether the contra-angle<br />

works.<br />

• Repeat the activation process<br />

and wait until the message<br />

“APEX LOCATOR ON” appears.<br />

• Alternatively, repeat the activation<br />

with the lip clip and file in the file clip,<br />

and wait until the message “APEX<br />

LOCATOR ON” appears.<br />

APEX LOCATOR IS<br />

OVERSENSITIVE – i.e.,<br />

it displays the apex too<br />

quickly or indicates too<br />

short a working length.<br />

• Short-circuit due to excess<br />

fluid in the pulp chamber<br />

(rinsing solution, saliva, blood).<br />

• Direct contact between file<br />

and gingiva/ gingival proliferation,<br />

i.e. in cases with fractured<br />

metal crowns.<br />

• Direct contact between file<br />

and metal restoration work<br />

(crowns, parapulpal pins,<br />

amalgam fillings).<br />

• Lateral root canal.<br />

• Immature canal with large<br />

apex.<br />

• Dry the access cavity using an air<br />

blower or a cotton ball.<br />

• If heavy bleeding is present, wait<br />

until bleeding has been successfully<br />

stopped.<br />

• For isolation:<br />

- Adequate tooth build-up<br />

- Electrocauterisation<br />

• Insert a <strong>dental</strong> dam.<br />

• Use the protective silicon cover on<br />

the contra-angle.<br />

• Carefully widen the access cavity;<br />

add flow <strong>com</strong>posites for isolation<br />

purposes if necessary.<br />

• Repeat length determination<br />

process.<br />

• Precise results not possible.