68<br />
Problem Possible Cause Solution<br />
en<br />
APEX STATUS LED<br />
does not switch to:<br />
red (stays off):<br />
green:<br />
Faulty connection between the<br />
lip clip cable and the unit.<br />
Defective lip clip cable.<br />
Lip clip and contra-angle file<br />
or lip clip and separate file clip<br />
connected too fast or incorrectly.<br />
• Contra-angle not functioning correctly.<br />
• Check whether the lip clip cable is<br />
inserted incorrectly or whether it is<br />
twisted or defective.<br />
• Clean the lip clip.<br />
• Check whether the measuring file<br />
has been correctly inserted into the<br />
contra-angle.<br />
• Check whether the contra-angle<br />
works.<br />
• Repeat the activation process<br />
and wait until the message<br />
“APEX LOCATOR ON” appears.<br />
• Alternatively, repeat the activation<br />
with the lip clip and file in the file clip,<br />
and wait until the message “APEX<br />
LOCATOR ON” appears.<br />
APEX LOCATOR IS<br />
OVERSENSITIVE – i.e.,<br />
it displays the apex too<br />
quickly or indicates too<br />
short a working length.<br />
• Short-circuit due to excess<br />
fluid in the pulp chamber<br />
(rinsing solution, saliva, blood).<br />
• Direct contact between file<br />
and gingiva/ gingival proliferation,<br />
i.e. in cases with fractured<br />
metal crowns.<br />
• Direct contact between file<br />
and metal restoration work<br />
(crowns, parapulpal pins,<br />
amalgam fillings).<br />
• Lateral root canal.<br />
• Immature canal with large<br />
apex.<br />
• Dry the access cavity using an air<br />
blower or a cotton ball.<br />
• If heavy bleeding is present, wait<br />
until bleeding has been successfully<br />
stopped.<br />
• For isolation:<br />
- Adequate tooth build-up<br />
- Electrocauterisation<br />
• Insert a <strong>dental</strong> dam.<br />
• Use the protective silicon cover on<br />
the contra-angle.<br />
• Carefully widen the access cavity;<br />
add flow <strong>com</strong>posites for isolation<br />
purposes if necessary.<br />
• Repeat length determination<br />
process.<br />
• Precise results not possible.