67<br />

Problem Possible Cause Solution<br />

THE BATTERY IS NOT<br />

WORKING PROPERLY<br />

• Despite observance of all<br />

precautionary measures, the<br />

battery discharges too quickly.<br />

• The device only works if the<br />

charger is plugged into the<br />

power supply – however not in<br />

battery mode.<br />

The battery could be damaged.<br />

Send the device to your service centre.<br />

en<br />

CALIBRATION ERROR 1<br />

An incorrectly-connected<br />

micromotor could disrupt the<br />

calibration process.<br />

Check whether the micromotor is<br />

correctly connected.<br />

CALIBRATION ERROR 2<br />

A VDW 6:1 contra-angle with<br />

excessive resistance can cause<br />

the calibration process to abort.<br />

• Check whether the contra-angle is<br />

malfunctioning.<br />

• Determine whether your contraangle<br />

care and maintenance routines<br />

are correct.<br />

• To rule out the possibility of the<br />

motor being defective, recalibrate the<br />

motor without the contra-angle.<br />

Rotary File<br />

Blocked in Canal<br />

• Incorrect file settings.<br />

• Too much pressure on the<br />

instrument.<br />

Switch to “backwards” ASR mode<br />

(red LED), restart the motor and carefully<br />

pull the file out.<br />

Reciprocating File<br />

Blocked in Canal<br />

• Too much pressure on the<br />

instrument.<br />

• File not cleaned often enough<br />

(see details in the RECIPROC ®<br />

user‘s manual).<br />

Remove the file using a pair of pliers<br />

by pulling it and carefully twisting it<br />

clockwise.<br />

The display reads<br />

APEX LOCATOR<br />

ERROR 3<br />

Apex locator cannot be switched<br />

on due to a malfunction.<br />

• Check whether the lip clip and/or file<br />

clip cables are inserted incorrectly or<br />

damaged.<br />

• Check whether you have acci<strong>dental</strong>ly<br />

connected the file in the contra-angle<br />

to the file clip instead of the lip clip.<br />

• Check whether the original VDW 6:1<br />

contra-angle is being used.<br />

• Check whether the new lip clip cable<br />

(included in your RECIPROC ® Upgrade<br />

Kit) and the original charger (both<br />

equipped with a ferrite ring) are being<br />

used.<br />