Problem Possible Cause Solution<br />

en<br />

The Apex Locator<br />

Will Not Turn On<br />

• One of the connection cables<br />

could be inserted improperly or<br />

defective.<br />

• Connection error between the<br />

cables and the unit.<br />

• Contact between the lip clip<br />

and either the contra-angle file<br />

or the file in the file clip either<br />

too short or faulty.<br />

• Check all plug connections and retry.<br />

• If you tried to activate the apex<br />

locator by connecting the lip clip to the<br />

file in the contra-angle, try again using<br />

the lip clip and the file in the file clip.<br />

If this successfully activates the apex<br />

locator, then the connection between<br />

the lip clip and the file in the contraangle<br />

is faulty.<br />

• Check the contact between the file<br />

and the lip clip.<br />

• If the results are still negative, this<br />

indicates that the contra-angle (only<br />

the original VDW 6:1 contra-angle may<br />

be used) is not transmitting signals<br />

correctly. The contra-angle needs to<br />

be replaced.<br />

• If the error persists following contact<br />

between the lip clip and the anglepiece<br />

file as well as between the lip<br />

clip and the file clip file, this could<br />

indicate a defective lip clip cable or<br />

a break in the measurement current<br />

circuit (file correctly inserted, file with<br />

electricity-conducting steel shaft).<br />

If the error cannot be fixed, switching<br />

the apex locator on and thus using it<br />

will not be possible.