Problem Possible Cause Solution<br />
en<br />
The Apex Locator<br />
Will Not Turn On<br />
• One of the connection cables<br />
could be inserted improperly or<br />
defective.<br />
• Connection error between the<br />
cables and the unit.<br />
• Contact between the lip clip<br />
and either the contra-angle file<br />
or the file in the file clip either<br />
too short or faulty.<br />
• Check all plug connections and retry.<br />
• If you tried to activate the apex<br />
locator by connecting the lip clip to the<br />
file in the contra-angle, try again using<br />
the lip clip and the file in the file clip.<br />
If this successfully activates the apex<br />
locator, then the connection between<br />
the lip clip and the file in the contraangle<br />
is faulty.<br />
• Check the contact between the file<br />
and the lip clip.<br />
• If the results are still negative, this<br />
indicates that the contra-angle (only<br />
the original VDW 6:1 contra-angle may<br />
be used) is not transmitting signals<br />
correctly. The contra-angle needs to<br />
be replaced.<br />
• If the error persists following contact<br />
between the lip clip and the anglepiece<br />
file as well as between the lip<br />
clip and the file clip file, this could<br />
indicate a defective lip clip cable or<br />
a break in the measurement current<br />
circuit (file correctly inserted, file with<br />
electricity-conducting steel shaft).<br />
If the error cannot be fixed, switching<br />
the apex locator on and thus using it<br />
will not be possible.