RF output<br />

RF Output<br />

1. RF ON/OFF Press [-ON/OFF) to turn the RF output signal on or off. RF OFF<br />

appears in the second line of the AMPLITUDE display when the<br />

output signal is off.<br />

The instrument turns off the output signal by swit<strong>ch</strong>ing in the<br />

maximum output attenuation (130 dB) and setting the vernier to<br />

its lowest setting. This results in approximately 170 dB of carrier<br />

isolation.<br />

2. ATTN HOLD Press [ATTN HOLD) to hold the step attenuator at its current setting.<br />

HOLD appears in the second line of the AMPLITUDE display when the<br />

attenuator hold function is on.<br />

1 b-54 Operation Reference<br />

When the attenuator hold function is on, amplitude adjustments are<br />

limited to the range of the instrument’s vernier. The vernier provides<br />

0.1 dB per step adjustment resolution across its specified 10 dB range.<br />

Vernier Ranges<br />

The following table, “10 dB Specified Vernier Ranges”, provides the<br />

upper and lower limits of ea<strong>ch</strong> vernier range. The instrument’s<br />

amplitude setting when you press the C-1 key determines<br />

whi<strong>ch</strong> vernier range is used.<br />

The vernier is allowed to over-range and under-range beyond the<br />

limits shown in the table when @iYlXKKj is selected. However,<br />

amplitude settings that exceed the limits may not provide output<br />

levels that are within the accuracy specifications of the instrument.