Time Base DAC<br />

All <strong>HP</strong> <strong>8648</strong>A/B/C/D<br />

A Manual Adjustment This automated adjustment can be replaced by performing the manual<br />

is Available Internal Reference Oscillator Adjustment instead.<br />

7-22 Adjustments<br />

Description This adjustment determines the proper DAC setting to a<strong>ch</strong>ieve a<br />

frequency reading of 1 GHz f 1 Hz. The DAC settings are then stored<br />

away.<br />

This program determines the proper DAC setting to a<strong>ch</strong>ieve a<br />

frequency reading of 1 GHz f 1 Hz. The DAC setting is then stored<br />

away.<br />

Required Test Spectrum Analyzer<br />

Equipment . Primary Frequency<br />

Procedure<br />

an61 2a2d<br />

<strong>HP</strong> 6646<br />

UNDER TEST<br />

RF Output<br />

Standard<br />

-<br />

I<br />

Input -<br />

5oQ<br />

Figure 7-22. Time Base DAC Test Setup<br />

1. Connect the equipment as shown above.<br />

2. Preset all of the equipment.<br />

3. Follow the instructions as they are displayed on the PC.<br />

I I<br />

PRIMARY<br />

FREQUENCY<br />

STANDARD<br />

10 MHz<br />

-2 to +iOdBm