RF Level Accuracy<br />

Performance Tkst<br />

Note This performance test is an adequate substitute for the automated<br />

Power Level Accuracy performance test. If the preferred Power Level<br />

Accuracy performance test is performed, this performance test is not<br />

required.<br />

Description This power accuracy verification procedure directly measures the full<br />

performance of the <strong>HP</strong> <strong>8648</strong>A.<br />

Connect the Test<br />

Equipment<br />

For the <strong>HP</strong> <strong>8648</strong>B/C/D, direct measurements can only be made over<br />

the full dynamic range up to 1300 MHz. The measuring receiver<br />

verifies the operation of the ALC and step attenuator over its full<br />

dynamic range, 13 dBm to -127 dBm for frequencies less than 1300<br />

MHz.<br />

Two power meters are used to verify operation at frequencies greater<br />

than 1300 MHz down to -60 dBm.<br />

Historically, testing of the <strong>HP</strong> <strong>8648</strong>B/C/D demonstrates that low-level<br />

performance at frequencies less than 1300 MHz is indicative of<br />

low-level performance at greater than 1300 MHz.<br />

iiii: 000000 ; ;;;; 0 SENSOR<br />

0 0 p 0000<br />

D 0 0 o-pop 0 0000<br />

lNPUT ’<br />

- 5 0 ”<br />

MEASURING RECEIVER<br />

8B888BBBF3<br />

00000000D<br />

000 0000~~<br />

UNDER TEST<br />

SENSOR MODULE<br />

Figure 8-16.<br />

Equipment Setup for the BP <strong>8648</strong>A and <strong>HP</strong> <strong>8648</strong>BICID 5 1300<br />

MHZ<br />

Configure the 1. Reset<br />

Measuring Receiver 2. RF power mode (tuned RF level mode for levels lower than<br />

-10 dBm).<br />

8-22 Performance Tests<br />

3. Connect the measuring receiver and <strong>HP</strong> <strong>8648</strong> timebases.<br />

4. Set to measure in dBm, press [LOG/LINJ<br />

Note Enter the power sensor’s calibration data into the measuring receiver<br />

and zero the sensor module. Refer to the measuring receiver’s<br />

operating manual.