HF Power Level<br />

Accuracy<br />

All <strong>HP</strong> <strong>8648</strong>A/B/C/D<br />

HF Power Level Accuracy<br />

Description This is the power level accuracy adjustment for the power range<br />

of 0 dBm to -130 dBm. The first step in this adjustment is to test<br />

the noise floor level to ensure at least 20 dB separation between the<br />

lowest measured power level and the spectrum analyzer noise floor. If<br />

the noise floor does not allow for 20 dB of separation at -130 dBm,<br />

then it is <strong>ch</strong>ecked at -120 dBm. If there is not 20 dB of separation<br />

at -120 dBm, the test will not be run. If, however, there is 20 dB of<br />

separation at -120 dBm, the -130 dBm correction factors will be set<br />

to the -120 dBm values.<br />

Required Test Power Meter<br />

Equipment Power Sensor<br />

Spectrum Analyzer<br />

Low Frequency Amplifier<br />

High Frequency Amplifier<br />

6 dB Attenuator<br />

Procedure<br />

POWER METER<br />

POWER SENSOR<br />

,<br />

<strong>HP</strong> <strong>8648</strong><br />

UNDER TEST<br />

Figure 7-26.<br />

HF Power Level Accuracy Test Setup<br />

for Power Levels > -10 dBm<br />

I Etput<br />

Adjustments 7-27