10 d 5 Specified Vernier Ranges<br />

Range Upper and Lower Limits<br />

1 + 10.0 dBm* to -5.9 dBm<br />

2 -6.0 dBm to - 15.9 dBm<br />

3 - 16.0 dBm to -25.9 dBm<br />

4 -26.0 dBm to -35.9 dBm<br />

5 -36.0 dBm to -45.9 dBm<br />

6 -46.0 dBm to -55.9 dBm<br />

7 -56.0 dBm to -65.9 dBm<br />

8 -66.0 dBm to -75.9 dBm<br />

9 -76.0 dBm to -85.9 dBm<br />

10 -86.0 dBm to -95.9 dBm<br />

11 -96.0 dBm to -105.9 dBm<br />

12 -106.0 dBm to -115.9 dBm<br />

13 - 116.0 dBm to - 127.0 dBm<br />

14 -127.1 dBm to -136.0 dBm<br />

* +13d Im for <strong>HP</strong> <strong>8648</strong>BICID<br />

RF Output<br />

3. RF OUTPUT The following table shows the reverse power protection for the RF<br />

output port of ea<strong>ch</strong> instrument model. When the instrument senses a<br />

reverse-power signal, it turns the RF output off, the step attenuator to<br />

maximum attenuation, and the vernier to its lowest setting.<br />

A message appears in the second line of the display when the reverse<br />

power protection has been activated.<br />

After you have removed the reverse-power signal from the RF output,<br />

press the (j-1 key to turn the output signal on again.<br />

Caution Applying a signal source to the RF output port that exceeds the power<br />

levels listed or maintaining a signal source at the RF output for an<br />

extended period of time may damage the instrument.<br />

Reverse Power Protection<br />

5 2000 MHz 50 watts into 50 ohms<br />

> 2000 MHz 25 watts into 50 ohms<br />

Operation Reference 1 b-55