5c-4 Service Error Messages<br />

512<br />

Description<br />

Cause<br />

What To Do<br />

513<br />

Description<br />

Cause<br />

What To Do<br />

515<br />

Description<br />

Cause<br />

What To Do<br />

516<br />

Description<br />

Cause<br />

What To Do<br />

Genericpathcalrestorefailure<br />

After detecting a failure in comparing calibration<br />

data between RAM and the calibration ROM, a data<br />

restore was attempted unsuccessfully.<br />

There is either a failure in writing data to RAM or<br />

in reading data from the calibration ROM.<br />

Replace the A3 assembly. The A3 assembly has a<br />

continuity failure. The calibration ROM is on the<br />

A3 assembly with the RAM.<br />

Output tune calrestorefailure<br />

After detecting a failure in comparing calibration<br />

data between RAM and the calibration ROM, a data<br />

restore was attempted unsuccessfully.<br />

There is either a failure in writing data to RAM or<br />

in reading data from the calibration ROM.<br />

The calibration ROM can be verified by replacing<br />

the A6 assembly. The calibration ROM is contained<br />

in the assembly.<br />

If the failure is still present after replacing the A6<br />

assembly, replace the A3 assembly, whi<strong>ch</strong> contains<br />

the RAM.<br />

FreqExtlevelcalrestorefailure<br />

After detecting a failure in comparing calibration<br />

data between RAM and the calibration ROM, a data<br />

restore was attempted unsuccessfully.<br />

There is either a failure in writing data to RAM or<br />

in reading data from the calibration ROM.<br />

The calibration ROM can be verified by replacing<br />

the A10 (freq ext) assembly. The calibration ROM is<br />

contained in the assembly.<br />

If the failure is still present after replacing the Al0<br />

assembly, replace the A3 assembly, whi<strong>ch</strong> contains<br />

the RAM.<br />

Configcalrestorefailure<br />

After detecting a failure in comparing calibration<br />

data between RAM and the calibration ROM, a data<br />

restore was attempted unsuccessfully.<br />

There is either a failure in writing data to RAM or<br />

in reading data from the calibration ROM.<br />

Replace the A3 assembly. The A3 assembly has a<br />

continuity failure. The calibration ROM is on the<br />

A3 assembly with the RAM.