Pager Encoder<br />

Timebase Accuracy<br />

Performance Test<br />

(Option 1EP Only)<br />

Connect the Test<br />

Equipment<br />

Pager Encoder Timebase Accuracy Performance Test (Option 1EP Only)<br />

<strong>HP</strong> a64aA UNDER TEST<br />

Configure the 1. Frequency Auto<br />

Frequency Counter 2. Input 1 MO<br />

I<br />

MOD<br />

1 NPUT/OUl ‘PUT<br />

UNIVERSAL COUNTER<br />

an64abc<br />

Figure S-20. Pager Encoder Timebase Accuracy Equipment Setup<br />

Configure the <strong>HP</strong> <strong>8648</strong> 1. Enter the pager encoder mode by pressing (FM) (ENCODER) twice.<br />

(If FM was the last active function, only press the IFhn) (ENCODER)<br />

key once.)<br />

2. Rotate the AMPLITUDE/ENCODER knob to set FORMAT to SERVICE.<br />

3. Press (JJ (NEXT) and rotate the AMPLITUDE/ENCODER knob to<br />

set CALIBRATIONFSK to 3200/2.<br />

4. Press @?i??Z] (START/STOP) to turn the encoder output on.<br />

Measure the 1. Record the test results and compare the results to the limits in the<br />

Frequency test record.<br />

Test Record <strong>HP</strong> <strong>8648</strong>A Option 1EP . . . . . . . . . . . .Table 8-17<br />