Performance lksts<br />

Calibration Cycle<br />

The procedures in this <strong>ch</strong>apter test the electrical performance of the<br />

signal generator. These tests do not require access to the interior of<br />

the instrument.<br />

This <strong>ch</strong>apter contains the following information:<br />

Required Test<br />

Equipment List<br />

A list of all the equipment required to<br />

perform these manual performance tests. For<br />

a comprehensive list of all test equipment<br />

required to perform these tests plus the<br />

manual adjustments, and the automated<br />

adjustments and performance tests, see<br />

Chapter 5, “Service”.<br />

Performance Tests The manual and automated performance tests<br />

required to verify the signal generator to its<br />

specifications. Most of these performance tests<br />

are performed manually, however, there is a<br />

single automated performance test.<br />

8<br />

The Bwer Level Accurucy performance test is<br />

the automated test. If this test is run, the RF<br />

L.evel Accuracy performance test, one of the<br />

manual tests, does not need to be performed.<br />

The description for this automated performance<br />

test is located after the descriptions for the<br />

manual performance tests. Refer to “Service<br />

Support Software” in Chapter 7 “Adjustments”<br />

for information on using the service support<br />

software that runs this performance test.<br />

Test Records A test record for ea<strong>ch</strong> model of the <strong>HP</strong> <strong>8648</strong> is<br />

located at the end of this <strong>ch</strong>apter. These test<br />

records are used to record the measurements<br />

for the manual performance tests. Keep these<br />

copies as masters and use a photocopy for ea<strong>ch</strong><br />

calibration.<br />

See Chapter 4, “Specifications, n for the specifications for ea<strong>ch</strong><br />

<strong>HP</strong> <strong>8648</strong> model.<br />

This instrument requires periodic verification of performance. Under<br />

normal use and environmental conditions, an instrument should be<br />

calibrated every 2 years. Normal use is defined to be about 2,000<br />

hours of use per year.<br />

Performance Tests 8-1