RF Level Accuracy Performance Test<br />

Configure the<br />

<strong>HP</strong> <strong>8648</strong>B/C/D<br />

Measure High Level<br />

Power Accuracy<br />

(5 13 dBm)<br />

Note<br />

Configure the<br />

<strong>HP</strong> <strong>8648</strong> Option 1EA<br />

1. Connect the power sensor to the <strong>HP</strong> <strong>8648</strong> RF OUTPUT connector.<br />

2. Set the amplitude; press CAMPLITUDE] (13) [dB(m).<br />

3. Turn the RF output on; press Cm].<br />

1. Enter the frequencies and power levels shown in the test record.<br />

Make sure the power sensor cal factors are entered correctly for<br />

ea<strong>ch</strong> frequency. Do not test below -20 dBm with this test setup.<br />

2. Record the test results and compare the results to the limits in the<br />

test record.<br />

<strong>HP</strong> <strong>8648</strong>B . . . . . . . . . . . . . . . . . . . . . . . . . . . .‘l%ble 8-38<br />

<strong>HP</strong> <strong>8648</strong>C . . . . . . . . . . . . . . . . . . . . . . . . . . . . Table 8-63<br />

<strong>HP</strong> <strong>8648</strong>D . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Table 8-88<br />

For <strong>HP</strong> <strong>8648</strong>B/C/D instruments without Option lEA, high power, skip<br />

ahead to “Configure the Power Meter”. All others, continue with the<br />

next step.<br />

1. Set the amplitude; press [AMPLITUDE) lloJ m.<br />

2. Turn the RF output on; press (RF).<br />

3. Connect the power sensor directly to the <strong>HP</strong> <strong>8648</strong> RF OUTPUT<br />

connector.<br />

Measure the High 1. Measure the RF power at ea<strong>ch</strong> frequency listed in Table 8-l and<br />

Power Level Accuracy record the measurement in the first row of the table.<br />

8-24 Performance Tests<br />

2. Connect the 10 dB attenuator between the <strong>HP</strong> <strong>8648</strong> RF OUTPUT<br />

connector and the power sensor.<br />

3. Measure the RF power at ea<strong>ch</strong> frequency listed in Table 8-l and<br />

record the measurement in the second row of the table.<br />

4. Calculate the actual attenuator value for ea<strong>ch</strong> frequency (the<br />

difference between the two measurements) and record it in the<br />

third row of the table.<br />

5. For ea<strong>ch</strong> frequency listed in Table 8-l set the <strong>HP</strong> <strong>8648</strong> amplitude<br />

to the level shown in the table and measure the RF power level<br />

through the 10 dB attenuator. Record the measurement in the fifth<br />

row of the table.<br />

6. Calculate the actual output power for ea<strong>ch</strong> frequency (the sum of<br />

the measured power and the attenuator value) and record it in the<br />

test record.