‘I&ble 8-1. High Power Level Accuracy Work ‘I&ble<br />

RF’ Level Accuracy Performance Test<br />

Amplitude Setting 20 20 20 20 19 17 15 13 13<br />

WW<br />

Measured<br />

Power<br />

Actual Output<br />

Power<br />

Configure the Power<br />

Meter<br />

Measure Low Level<br />

Power Accuracy<br />

(< -20 dBm)<br />

Test Record<br />

7. Record the test results and compare the results to the limits in the<br />

test record.<br />

1.<br />

2.<br />

3.<br />

1.<br />

2.<br />

3.<br />

<strong>HP</strong> <strong>8648</strong>B . . . . . . . . . . . . . . . . . . . . . . . . . . ‘fable 8-39<br />

<strong>HP</strong> <strong>8648</strong>B with Options 1EA and lE6 . . . . . . . . . Table 8-40<br />

<strong>HP</strong> <strong>8648</strong>C . . . . . . . . . . . . . . . . . . . . . . . . . ‘Ihble 8-64<br />

<strong>HP</strong> <strong>8648</strong>C with Options 1EA and lE6 . . . . . . . . .Table 8-65<br />

<strong>HP</strong> <strong>8648</strong>D . . . . . . . . . . . . . . . . . . . . . . . . . . . . .‘lhble 8-89<br />

<strong>HP</strong> <strong>8648</strong>D with Options 1EA and lE6 . . . . . . . . . . . . . . . . . . Table 8-90<br />

Connect the <strong>HP</strong> 8481D Option H70 power sensor to the power<br />

meter.<br />

Connect the power sensor through the <strong>HP</strong> 11708A 30 dB<br />

attenuator to the power meter Power Ref output.<br />

Zero and calibrate the power meter.<br />

Connect the power sensor to the <strong>HP</strong> <strong>8648</strong> RF OUTPUT connector.<br />

Enter the frequencies and power levels shown in the test record.<br />

Make sure the power sensor cal factors are entered correctly for<br />

ea<strong>ch</strong> frequency. Do not test at levels > -20 dBm with this test<br />

setup.<br />

Record the test results and compare the results to the limits in the<br />

test record.<br />

<strong>HP</strong> <strong>8648</strong>B ................................................... Table 8-41<br />

<strong>HP</strong> <strong>8648</strong>C ................................................... Table 8-66<br />

<strong>HP</strong> <strong>8648</strong>D .................................................. .l%.ble 8-91<br />

Performance Tests 8-25

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!