6-10 Chapter 6 Fault Diagnosis! "I<br />

Fitting a New LaB, roament<br />

The new emitter should be fitted, accurately, according to Section 2.2.4 Changing<br />

the Filament.<br />

Running Up an LaB6 Filament<br />

ALaB6 emitter should also be run up using the instrument's automatic procedure<br />

(selcting Beam On from the gun control box on the toolbar). The procedure is<br />

designed to maximise the filament life. Similarly the Beam Off procedure<br />

should always be used to run the gun down.<br />

~<br />

I<br />

~<br />

The only exception to this is when a Kimball LaB6 emitter is installed for the first<br />

time. To ensure that the Kimball emitter is saturated at the correct filament<br />

current (1.8-2.0A) it should be run up slowly using the Manual option and the<br />

Gun Set Up panel. Once the filament has been saturated, the automatic routines<br />

can be used safely.<br />

The most imrortant factor of all in maximising LaB6 life is to ensure that the<br />

emitter has been fully cooled down before exposing it to air, as otherwise, the<br />

LaB6 crystal is liable to be severely damaged, or even destroyed.<br />

LaB, Beam Current.<br />

The beam current is fixed for LaB6 operation, and not user changeable.<br />

~<br />

I;i.'<br />

rI<br />

t"l<br />

6.4. LEO Error Messages<br />

There are many reasons why error messages may appears. With each message<br />

on-line help is available. In addition an alphabetical list of all error messages is<br />

included below.<br />

6.4.1. E"or Messages<br />

ACQUIRE COMMAND ERROR<br />

This error is given when a signal acquisition command fails.<br />

Code numbers greater than 40 indicate the failure of an internal consistency<br />

check. Contact Leica<br />

Code numbers less than 40 indicate that there is a failure in the communication<br />

to the DSP subsystem.<br />

~<br />

I<br />

fI<br />

rn<br />

<strong>Stereo</strong>scan <strong>440</strong> Operator Manual<br />

~<br />