" Chapter 3 Advanced Operation 3-13<br />

Figure 3.8(b) Taken at a slow scan speed<br />

The abow shows a sample of Ceramic<br />

3.5. Specimen Preparation<br />

Specimens which are studied in the SEM can be divided into two main<br />

categories, namely conductors and non-conductors. Factors to consider during<br />

specimen preparation are given below:-<br />

The size of the specimen must be reduced (if necessary) to fit the<br />

available specimen holders and specimen stages, but there is often an<br />

advantage in selecting an even smaller sample size for ease of<br />

roservation.<br />

2 The object must be able to withstand being in the high vacuum of the<br />

SEM; it must not change its shape, and it must not outgas. A cold stage<br />

may help here.<br />

3, It should be clean, ie free of dust, oils and greases. (Their presence can<br />

lead to charging and contamination effects.)<br />

4. It may be treated to improve the secondary yield if this is low, eg coating<br />

of the sample with gold.<br />

s. Any disturbance in the surface structure caused by preparation<br />

procedures should give rise to surface details which are too fme to be<br />

resolved at the magnification used to record the images.<br />

<strong>Stereo</strong>scan <strong>440</strong> Operator Manual