FIgUre 1.10 Image of First Peak Setting<br />

of)<br />

The filament current may be varied by using the associated scroll bar within the<br />

Gun Alignment panel. either by SLOWL Y dragging the slider in the appropriate<br />

direction. or by clicking over the arrow at the appropriate end of the scroll bar.<br />

This should be done carefully and slowly, otherwise a blown filament may<br />

result.<br />

Alignment<br />

For the best performance of the microscope the electron beam should pass down<br />

the column along the electron optical axis, and any aperture in use should also be<br />

centred about this axis. As no column can be perfect, various adjustments are<br />

provided to compensate for this.<br />

Final Aperture Alignment<br />

The centring of the final aperture can easily be checked by having a recognisable<br />

feature at the centre of screen and then taking the focus through focus and back<br />

again. When aligned. the image will only rotate and not shift laterally. If the<br />

aperture is mis-aligned the image will move backwards and forwards across the<br />

screen in a slight arc. If the aperture is mis-aligned it will be necessary to move<br />

the position of the aperture using the X and Y micrometers on the side of the<br />

column until the image ceases to shift laterally.<br />

<strong>Stereo</strong>scan <strong>440</strong> Operator Manual