Stage <strong>Scan</strong> Invalid<br />

This indicates if the current stage scan pattern is invalid or not.<br />

The scan will be invalid if the combination of scan parameters would take the<br />

stage beyond the user defined stage limits (in X, Y or Z).<br />

Stage <strong>Scan</strong>ning<br />

This parameter indicates the current state of stage scanning:-<br />

Idle Stage scan has not been started<br />

At Field Stage is positioned at a field in the scan pattern<br />

Moving The stage is in motion as a result of a Next field Command<br />

At End The scan pattern has been completed<br />

Stage XY +Z<br />

This state is used when tilt compensation is operative.<br />

If the checkbox is empty then no Z movement is made.<br />

If the checkbox is marked then movement in Z is made relative to the startZ<br />

when movement is made in the tilt direction.<br />

Backlash<br />

If this box is checked backlash correction will be perfonned on each stage step.<br />

Start Here<br />

Executing this command will set the start co-ords to the current stage position.<br />

End Here<br />

Executing this command will set the End co-ords to the current stage position.<br />

Calc Fields<br />

This function performs the following calculation.<br />

The field of view is determined from the current magnification (i.e. the physical<br />

dimensions on the specimen of the image in nonnal mode).<br />

The field of view is used with the field overlap to determine the field size.<br />

Using the Start co-ord. the End Co-ord and the field size the number of fields<br />

is determined.<br />

Select scan area<br />

This function is used to set the Start co-ord and End Co-ord when the whole of<br />

the area to be scanned is visible at a suitable low magnification.<br />