2.3 laB,<br />

~<br />

As soon as a LaB6 filament has been fitted into the gun it is advisable to set the<br />

software to recognise that a LaB6 is present. This can be done by calling up the<br />

Gun Set Up panel from the kV adjustment icon. Click on the Filament Type =<br />

box from within the Gun selections, to call up a list of different filaments. Select<br />

the appropriate filament type, probably this will be Denka M3. Also, if<br />

appropriate the New Filament box can be selected, see below. Setting the<br />

fIlament type will prevent the filament from being turned on when the vacuum<br />

conditions are incoITect.<br />