E-32 Appendix E Qptions<br />

.<br />

.<br />

High User Limit<br />

High Inner Limit<br />

High outer Limit<br />

The current limit state may be displayed in the Slams window by selecting the<br />

parameters X limit hit. Y Limit hit and Z limit hit.<br />

A user limit is a software limit. They may be selected in the status window (Stage<br />

High X, Stage Low X, Stage High Y, Stage Low Y, Stage High Z and Stage Low<br />

Z). To change a user limit click on the required limit and enter the value. Note<br />

the limits may be set by a macro.<br />

If a user limit is hit a warning is given and the stage slowed to a halt. Position is<br />

not lost.<br />

If an Inner limit is hit the motor is slowed to a halt and the axis may only be<br />

moved away from the limit.<br />

If an outer limit is bit the motors are stopped and may only be moved manually.<br />

Position is lost.<br />

If the touch alarm is detected a buzzer is sounded, an error is given, position is<br />

lost and the motors must be moved manually until the buzzer is silent.<br />

Controls<br />

Stage Initialisation<br />

The Stage lnitialisation panel is used to define the actual position and is<br />

displayed by Top Menu - StageNac - Stage lnit. On initial installation, or if an<br />

axis outer limit limit has been hit, or if the touch alarm has occurred the stage<br />

position is unknown. The Stage lnitialisation panel is used to define the actual<br />

position and is displayed by Top Menu - StageNac - Stage lnit.<br />