Running the Filament Up<br />

Standard run-up macros for Denka and Kimball LaB6 filaments are provided by<br />

Leica. These will set the EHT to 5kV and the filament current to a value<br />

normally below the saturation point for the appropriate filament type. It is<br />

recommended that these are used initially when running with LaB6 and then<br />

when the desired operating conditions are achieved that additional run-up macros<br />

are created for future use.<br />

~<br />

Kimball<br />

To select the appropriate macro select File from the top menu line followed by<br />

Load state. This produced a list of application macros from which the<br />

appropriate run-up macro can be selected. Select DenkaS.app for Denka M3<br />

filaments and KimbS.app for Kimball Physics filaments.<br />

filaments run with a much lower heating current so the selection of<br />

an incorrect macro could result in the filament being destroyed.<br />

Once Open has been selected the macro will be loaded and run up of the filament<br />

will commence. If there is a conflict between the conditions in the macro and<br />

those currently set on the system then the macro will halt and a warning will be<br />

given and the Operator will be asked whether or not to proceed. It is advisable to<br />

select No and to investigate the reason for the conflict. Once this has been<br />

corrected then the macro can be loaded and run again.<br />

Once the run up sequence is completed it is advisable to check for correct<br />

filament saturation and beam alignment. Use the Gun Align panel to select<br />

Emiaion Image mode and adjust the filament cuuent to produce the desired<br />

emission image (see Figures 9,10,11,12).<br />