4. Select SCM Image from the Specimen Current panel or from the<br />

Status! menu, set Brightness to 50% and adjust Contrast until an<br />

image is seen. Then adjust Brightness and Contrast to obtain the<br />

required image.<br />

At low probe currents i.e. on SCM ranges lower than loA full scale, the image<br />

can become degraded by mains field effects and other forms of interference. Also,<br />

on these ranges, the signal bandwidth is very low. The use of long scan times will<br />

overcome the bandwidth restrictions and will also help to reduce interference<br />

effects, but image quality may not be perfect under these conditions.<br />

Measure Mode<br />

Because of the way in which the SCM is configured, selecting Measure mode<br />

increases the Image mode signal level. If you wish to measure the specimen<br />

current in a displayed specimen image, Freeze the image before proceeding.<br />

Select SCM Measure from the Specimen Current panel or from the<br />

Status! menu.<br />

2. The specimen current measured will be displayed beside the Specimen I<br />

legend in the Specimen Current panel or in the Status! menu.<br />

3. To measure the current at one particular point on the specimen, select<br />

Spot from the <strong>Scan</strong>ning menu, click on the cross and drag it to the point<br />

of interest.<br />

Specification<br />

Measurement mode<br />

Measurement Range<br />

Bandwidth<br />

Resolution<br />

Accuracy<br />

Residual Error<br />

lOpA to l00~ in 12 ranges (auto ranging)<br />

5Hz<br />

IpA<br />

:t 2% of Full Scale<br />

< IOpA<br />

Image Mode<br />

Image Range (Full)<br />

Image Range (Useful) *<br />

Bandwidth<br />

3pA to IOJiA in 13 ranges (selected by Contrast)<br />

InA to IOJiA<br />

6kHz to 2MHz (see Table on next page)<br />

<strong>Stereo</strong>scan <strong>440</strong> Operator Manual