Each zone has its individual detector assignment, so in quad mode as many as<br />

four different detectors could be displayed. The detector controlled by the signal<br />

level functions is detemlined by the current zone, although if that detector is<br />

assigned to more than one zone then more than one part of the quad image may<br />

be affected.<br />

~<br />

It is possible to set up different auto signal level controls for different<br />

zones. If there is a conflict between two zones which have the same detector<br />

assigned to them, then the first zone with that detector will take precedence.<br />

Collector Bias<br />

The Collector Bias can be assigned to mouse control from the option in the<br />

Detectors menu. It can be varied in the range -250V to +4OOV. For nonnal<br />

operation of the secondary electron detector this value should be set to maximum.<br />

Four Quadrant BSD<br />

The four quadrant back scattered detector has a configuration panel dedicated to<br />

it, which can be selected from the Detectors menu. Each quadrant of the detector<br />

can be in one of three states (Off, Normal or Invert). When normal the signal<br />

from that quadrant is added to the total signal, when inverted the signal is<br />

subtracted. Clicking on the quadrant fields allows each one to be configured as<br />

desired.<br />

Qf<br />

Notmal<br />

02<br />

NotmaJ<br />

01<br />

kIVett<br />

02<br />

krveIt<br />

03<br />

Nom8J<br />

04<br />

NomIaI<br />

03<br />

Normal<br />

04<br />

NomIai<br />

Cof1f)OSIkJn<br />

T~y<br />

FIgUre 2.5<br />

Two predefined configurations of quadrants are available to set the detector to<br />

give best perfonnance for composition (Compo), that is atomic number contrast,<br />

or surface topography (Topo). Selecting these predefined configurations will set<br />

up the detector quadrants according to Figure 2.5.<br />

<strong>Stereo</strong>scan <strong>440</strong> Operator Manual