Chapter 2 Operation 2-15<br />

B SIstems without the Modified Gun Ali2D Modules<br />

Build Standard 3/9 or ealier<br />

(i) Software Version 2.01 and earlier<br />

Use an Emission Zoom Factor of 1 to display an emission image<br />

of the correct size.<br />

(ii) Software Version 2.02 and later<br />

Use an Emission Zoom Factor of 1 to display an emission image<br />

of the correct size (will be slightly bigger).<br />

3. Select Optibeam mode depth in the Gun Align panel.<br />

4.<br />

Select Gun Shift.<br />

,. Defocus the image.<br />

6.<br />

Adjust the magnification until an oval spot is clearly visible, - lOOx.<br />

7.<br />

8.<br />

Centre the spot using the crosshairs if required.<br />

Undo Optibeam depth mode.<br />

9. Select Gun Tilt.<br />

10. Centre the emission image. if required<br />

Select normal image.<br />

12. Select reduced raster<br />

13. Select the gun alignment icon.<br />

14. Adjust for maximum brightness in X and Y directions.<br />

15. Press the normal icon to resume normal imaging.<br />

If the image moves laterally when you adjust the focus, it may be necessary to<br />

adjust the aperture position. If so, consult Section 2.3.3 Adjusting the Aperture.<br />