Chapter 2 Operation<br />

2-23<br />

Manual<br />

The signal level can be adjusted using the brightness and contrast controls<br />

assigned to the mouse with the signal level icon. If more than one zone is<br />

displayed, then the detector controlled is the detector assigned to the current zone<br />

(which is indicated by the anchor symbol). The concept of zones will be described<br />

in the Section 2.4.2 <strong>Scan</strong>ning Modes.<br />

Auto Signal Level<br />

The Signal Adjust panel provides the controls for the auto signal level functions.<br />

Brightness and contrast can also be controlled directly from this panel using the<br />

slider bars labelled accordingly.<br />

There are two types of auto signal level function which each operate in two ways.<br />

The auto signal level function will either control the Mean signal level or the<br />

Peak signal level. The mean signal level is defined as being the average<br />

brightness of all the pixels in the image, while the peak level is defined as the<br />

brightness of the brightest part of the image.<br />

Peak and mean signal level controls each operate in two ways. The Find function<br />

is a single shot activation which sets the peak or mean to the level defined in the<br />

associated Auto Video slider bar. The Keep function activates the auto video<br />

function to maintain the video level (peak or mean) of the image at the time of<br />

selection.<br />

Typically the Find function will be used to set the required video level if the<br />

image is a long way from the required brightness. The Keep function is useful<br />

when the content of the image is likely to change, for example if the stage is<br />

being moved across a specimen with distinct light and dark phases, or if the<br />

probe current is being adjusted.<br />

Auto video conttol can be disabled at any time by selecting the Auto Video Off<br />

button.<br />

2.4.2. Using <strong>Scan</strong> Modes<br />

Before considering the different scan modes available, it is necessary to define<br />

some tenDS which will simplify the whole issue of scan modes, signals and<br />

Windows.<br />

<strong>Stereo</strong>scan <strong>440</strong> Operator Manual