1. Admit air to the gun. Open the gun.<br />

2. Lift out die anode.<br />

3. Fit the anode spacer on the top of the gun align coils. Take care, this is a<br />

high precision fit.<br />

4 Fit the anode back. on top of the spacer. This is also a good fit.<br />

~<br />

This<br />

S. Select Low kV anode in the Gun Setup panel.<br />

automatically restricts the acceleration vollage to 3kV.<br />

6 Close the gun. pump down. select the desired low voltage and obtain an<br />

image.<br />

The Slereoscan <strong>440</strong> should not be used at an acceleration voltage higher than<br />

3k V with the anooe spacer fitted. It is important to ensure that the state of the<br />

anode spacer p~ter reflects the state of the machine.<br />

2.2.6. Getting a Vacuum<br />

If the vacuum ready state is not achieved after several minutes of pumping, the<br />

following guidelines provide an ordered approach to diagnosing the problem.<br />

Investigating a vacuum leak. few simple rules:-<br />

Before looking for a vacuum leak give the vacuum system time to pump. If<br />

you have just changed a speci~n in a very humid atmosphere the pump<br />

down may take longer. If the specimen is wet it may take several minutes<br />

(up to I hour with a very large. very wet specimen has been known.) If the<br />

stage has been left open for long periods it will take longer to pump. If the<br />

column or chamber have been washed with solvents then it may take several<br />

hours to achieve a good vacuum.<br />

2. Vacuum leaks rarely happen by themselves; they are often caused. If you<br />

have just done anything to the column or chamber then that is the most likely<br />

cause of the leak. If the s~ has just been changed. check the stage<br />

door '0' ring. Similarly if the filament has been changed check the gun '0'<br />

ring.<br />

<strong>Stereo</strong>scan <strong>440</strong> Operator Manual