Chapter 4 Calibration Procedures<br />

Relay Verification<br />

Tests 9 - 10:<br />

Open all channels on the module by performing a Factory Reset.<br />

Configure Channel 10 (module in slot 200) as follows: 4-wire ohms, 1 k<br />

range, and 5½ digits.<br />

Enable reading monitoring by pressing on the selected channel (or<br />

use the ROUTe:MON command). Record the 4-wire ohms measurements<br />

from the external DMM in the following table.<br />

External DMM Ohmmeter Connections<br />

Test<br />

#<br />

Channel<br />

Configured HI LO<br />

HI<br />

Sense<br />

LO<br />

Sense<br />

Measured Value<br />

Test<br />

Limit<br />

Relay<br />

Measured<br />

9 Ch 10 P2 P2 P3 P3 ________Ohms — —<br />

10 Subtract (Test 9 - Test 3) ________Ohms 2.00 K423<br />

Tests 11- 33:<br />

Open all channels on the module by performing a Factory Reset. For<br />

each test, close only the channel shown in the “Channel Closed” column<br />

below (module in slot 200). Turn the Monitor Mode “off” and select<br />

“Banks Joined” from the Advanced menu. Record the 4-wire ohms<br />

measurements from the external DMM in the following table.<br />

External DMM Ohmmeter Connections<br />

Test<br />

#<br />

Channel<br />

Closed*<br />

HI LO HI<br />

Sense<br />

LO<br />

Sense<br />

Measured Value<br />

Test<br />

Limit<br />

Relay<br />

Measured<br />

11 Ch1 P2 P2 P1 P1 ________Ohms 2.00 K401<br />

12 Ch 2 P2 P2 P1 P1 ________Ohms 2.00 K402<br />

13 Ch 3 P2 P2 P1 P1 ________Ohms 2.00 K403<br />

14 Ch 4 P2 P2 P1 P1 ________Ohms 2.00 K404<br />

15 Ch 5 P2 P2 P1 P1 ________Ohms 2.00 K405<br />

16 Ch 6 P2 P2 P1 P1 ________Ohms 2.00 K406<br />

17 Ch 7 P2 P2 P1 P1 ________Ohms 2.00 K407<br />

*Only the channel currently under test should be closed at one time. All other channels should be open.<br />