Chapter 4 Calibration Procedures<br />
Internal DMM Verification Tests<br />
AC CURRENT Gain Verification Test<br />
Configuration: AC Current<br />
LF 3 HZ:SLOW (in the Advanced menu)<br />
1 Make sure you have read “Test Considerations” on page 81.<br />
2 Select Channel 221, set the AC CURRENT function and the 3 Hz input<br />
filter. With the slow filter selected, each measurement takes 7 seconds to<br />
complete. Before executing each test, you must press to enable<br />
reading monitoring on the selected channel (or use the ROUTe:MON<br />
command from the remote interface).<br />
3 Select each range in the order shown below. Compare measurement<br />
results to the appropriate test limits shown in the table. (Be certain to<br />
allow for appropriate source settling.)<br />
1 kHz<br />
Input Range Quick Error from Nominal<br />
Check<br />
Current Frequency 24 hour 90 day 1 year<br />
10 mA [1]<br />
10 mA<br />
± 14 A ± 14 A ± 14 A<br />
10 mA<br />
1 A [1] 1 kHz 1 A<br />
± 1.41 mA ± 1.41 mA ± 1.41 mA<br />
1 kHz 1 A<br />
± 1.4 mA ± 1.4 mA ± 1.4 mA<br />
100 mA [1] 1 kHz 100 mA Q ± 600 A ± 600 A ± 600 A<br />
[1] Verify only, no adjustment.<br />
4<br />
Frequency Gain Verification Test<br />
Configuration: Frequency<br />
6½ digits<br />
1 Make sure you have read “Test Considerations” on page 81.<br />
2 Select Channel 210, select the FREQUENCY function and set 6½ digits.<br />
3 Select each range in the order shown below. Compare measurement<br />
results to the appropriate test limits shown in the table. (Be certain to<br />
allow for appropriate source settling.)<br />
Input Range Quick Error from Nominal<br />
Check<br />
Voltage Frequency 24 hour 90 day 1 year<br />
10 mV [1]<br />
1 V<br />
100 Hz<br />
100 kHz<br />
100 mV<br />
1 V Q<br />
± 0.06 Hz<br />
± 6 Hz<br />
± 0.1 Hz<br />
± 10 Hz<br />
± 0.1 Hz<br />
± 10 Hz<br />
[1] Verify only, No adjustment. For this test, isolate the calibrator’s output from earth ground.<br />
Q: Quick performance verification test points.<br />