Chapter 3 Front-Panel Overview<br />

To Unsecure for Calibration<br />

To Unsecure for Calibration<br />

You can unsecure the instrument either from the front panel or over the<br />

remote interface. The instrument is secured when shipped from the<br />

factory and the security code is set to “HP034970” or “AT034972”,<br />

depending on the product number.<br />

• Once you enter a security code, that code must be used for both frontpanel<br />

and remote operation. For example if you secure the<br />

instrument from the front panel, you must use that same code to<br />

unsecure it from the remote interface.<br />

• Press to enter the Utility menu.<br />

When you first enter the Utility menu, the calibration entries toggle<br />

between CAL SECURED and UNSECURE CAL. To unsecure the<br />

instrument, select UNSECURE CAL and press . After entering the<br />

correct security code, press again. When you return to the menu,<br />

you will see new choices CAL UNSECURED and SECURE CAL.<br />

Note: If you enter the wrong secure code, NO MATCH is displayed and a<br />

new choice, EXIT, is shown.<br />

60