Chapter 4 Calibration Procedures<br />
Internal DMM Verification Tests<br />
...Continued from previous page<br />
3 Select each function and range in the order shown in the table below.<br />
Before executing each test, you must press to enable reading<br />
monitoring on the selected channel (or use the ROUTe:MON command<br />
from the remote interface). Compare measurement results to the<br />
appropriate test limits shown in the table (see page 84).<br />
Input<br />
Open<br />
Open<br />
Open<br />
Input<br />
Short<br />
Short<br />
Short<br />
Short<br />
Short<br />
Short<br />
Short<br />
Short<br />
Short<br />
Short<br />
Short<br />
Short<br />
Channel 221<br />
Quick<br />
Error from Nominal<br />
Function [1] Range Check 24 hour 90 day 1 year<br />
DC Current 10 mA<br />
± 1 A ± 2 A ± 2 A<br />
100 mA<br />
Q ± 4 ± 5 ± 5 <br />
1A<br />
± 60 A ± 100 A ± 100 A<br />
Channel 209<br />
Quick<br />
Error from Nominal<br />
Function [1] Range Check 24 hour 90 day 1 year<br />
DC Volts<br />
2-Wire Ohms [2]<br />
and<br />
4-Wire Ohms<br />
100 mV<br />
1V<br />
10V<br />
100V<br />
300V<br />
100 <br />
1 k<br />
10 k<br />
100 k<br />
1 M<br />
10 M<br />
100 M<br />
Q<br />
Q<br />
± 3.5 V<br />
± 6 V<br />
± 40 V<br />
± 600 V<br />
± 6 mV<br />
± 3.5 m<br />
± 6 m<br />
± 50 m<br />
± 500 m<br />
± 10 <br />
± 100<br />
± 10 k<br />
± 4 V<br />
± 7 V<br />
± 50 V<br />
± 600 V<br />
± 9 mV<br />
± 4 m<br />
± 10 m<br />
± 100 m<br />
± 1 <br />
± 10 <br />
± 100<br />
± 10 k<br />
± 4 V<br />
± 7 V<br />
± 50 V<br />
± 600 V<br />
± 9 mV<br />
± 4 m<br />
± 10 m<br />
± 100 m<br />
± 1 <br />
± 10 <br />
± 100<br />
± 10 k<br />
[1] Select 6½ digit resolution.<br />
[2] for 2-wire ohms, an additional 4 of error must be added.<br />
Q: Quick performance verification test points.<br />
Note: Zero offset calibration using a multifunction calibrator is NOT<br />
recommended. The calibrator and cabling offset can be large and<br />
unstable causing poor offset calibration of the internal DMM.<br />
86