Chapter 4 Calibration Procedures<br />

Internal DMM Adjustments<br />

Valid Gain Adjustment Input Values<br />

Gain adjustment can be accomplished using the following input values.<br />

DC VOLTS<br />

Function Range Valid Calibration Input<br />

Values<br />

100 mV to 100 V<br />

300 V<br />

0.9 to 1.1 x Full Scale<br />

250 V to 303 V<br />

OHMS, OHMS 4W<br />

100 to 10 M<br />

0.9 to 1.1 x Full Scale<br />

DC CURRENT<br />

10 mA to 1 A<br />

0.9 to 1.1 x Full Scale<br />

AC VOLTS [1]<br />

AC CURRENT<br />

10 mV to 100 V<br />

300 V<br />

1 A<br />

0.9 to 1.1 x Full Scale<br />

95 V to 303 V<br />

9 mA to 11 mA<br />

4<br />

Frequency<br />

Any<br />

Any Input >100 mV rms,<br />

1 kHz - 100 kHz<br />

[1] Valid frequencies are as follows: 1 kHz ± 10% for the 1 kHz calibration,<br />

45 kHz - 100 kHz for the 50 kHz calibration,<br />

and 10 Hz ± 10% for the 10 Hz calibration.<br />

93