Chapter 5 Theory of Operation<br />
Internal DMM<br />
Input<br />
Unless otherwise noted, components in this discussion are located on the<br />
A4 circuit assembly (34970-66504).<br />
The purpose of the Input section is to connect the Input HI terminal to the<br />
various measuring functions. This is accomplished through K102, K103,<br />
and K104. Additionally, connections are made for the 4-wire ohms<br />
HI Sense and LO Sense inputs. Shunt selection (ranging) and voltage<br />
sensing are also performed for the current function. The table below<br />
shows the state of each relay for each measuring function. All relay coils<br />
are driven from U150.<br />
Functional K102 K103 K104 Sense at:<br />
0.1 V - 10 VDC<br />
100 V - 300 VDC<br />
2-Wire Ohms<br />
4-Wire Ohms<br />
AC Voltage<br />
Frequency/Period<br />
3 A, 1A DCI<br />
100 mA, 10 mA DCI<br />
1A AC I<br />
Set<br />
Set<br />
Reset<br />
Reset<br />
Set<br />
Set<br />
Reset<br />
Set<br />
Reset<br />
Set<br />
Reset<br />
Set<br />
Set<br />
Reset<br />
Reset<br />
Reset<br />
Reset<br />
Reset<br />
Set [1]<br />
Set<br />
Reset [2]<br />
Reset [2]<br />
Reset<br />
Reset<br />
Reset<br />
Reset<br />
Set<br />
U101-5<br />
U102-12<br />
U101-5 [3]<br />
AC_IN<br />
AC_IN<br />
U101-10<br />
U101-10<br />
AC_IN<br />
[1] K104 will be reset when input resistance is selected to >10,000 M through the menu.<br />
[2] K104 will be set for the 100 M range.<br />
[3] Configurations shown are for the current source output (HI) terminal. The measurement<br />
sense is accomplished through the Sense HI / Sense LO terminals.<br />
140