Chapter 5 Theory of Operation<br />
Front Panel<br />
Front Panel<br />
Unless otherwise noted, components in this discussion are located on the<br />
A2 circuit assembly (34970-66502).<br />
The front-panel circuits consist of vacuum fluorescent display control,<br />
display high voltage drivers, and keyboard scanning. Communication<br />
between the front panel and floating logic circuits is accomplished<br />
4<br />
through a 2-wire bi-directional serial interface. The front-panel logic<br />
operates from -13 volts (logic 1) and -18 volts (logic 0). The two serial<br />
communication signals are level shifted by comparator U6 from the<br />
floating logic 0 V to 5 V levels to the -18 V to -13 V levels present on the<br />
front panel assembly. The front panel logic high supply (-13 volts) is<br />
produced from the -18 volt supply by voltage regulator U7.<br />
Display anode and grid voltages are +18 volts for an on segment and -18<br />
volts for an off segment. The -11 V cathode bias for the display is<br />
provided by the main pc boards filament winding center tap bias circuit<br />
A1CR108, A1R106, and A1C106.<br />
Keyboard scanning is accomplished through a conventional scanned rowcolumn<br />
key matrix. Keys are scanned by outputting data at<br />
microprocessor U1 port pins P0.0 through P0.4 to poll each key column<br />
for a key press. Column read-back data are read by the microprocessor at<br />
port pins P2.0 through P2.3 for decoding and communication to the<br />
floating logic circuits. Rotary knob quadrature inputs are read directly<br />
by the microprocessor port pins P2.6 and P2.7.<br />
The standby power switch, S19, provides a low true signal to main<br />
controller A1U205. In turn, A1U205 takes actions that either place the<br />
instrument in the “standby mode” or “on” mode. In “standby”, both the<br />
filament supply to the front panel and the +5V_BP supply to the<br />
backplane, rear panel interfaces, and fan are turned off.<br />
5<br />
137