Chapter 5 Theory of Operation<br />

Floating Logic<br />

Floating Logic<br />

Unless otherwise noted, components in this discussion are located on the<br />

A1 circuit assembly (34970-66501).<br />

The floating common logic controls the operation of the entire<br />

instrument. All measurement control and remote interface command<br />

interpretation is performed in the main controller, U205. The front panel<br />

controller, the I/O controller, and all of the plug-in module controllers,<br />

act as slaves to U205. The floating common logic is comprised of the main<br />

controller U205, custom ASIC U209, calibration memory U201, 12 MHz<br />

clock oscillator U204, and microprocessor supervisor U104.<br />

The microprocessor supervisor U104 performs the following functions:<br />

• Monitors the regulated 5V floating supply and generates the reset<br />

signal for main controller U205 when the voltage drops below<br />

operating levels.<br />

• Monitors the unregulated side of the 5V floating supply in order to<br />

generate an early warning signal (PWRFAIL) when power is lost.<br />

• Provides automatic switch over to the battery BT101 for the +5V_NV<br />

supply when the 5V_FLT supply drops below the battery voltage.<br />

• Blocks the main controller’s write signal (WR_N) while the 5V_FLT<br />

supply is below operating level.<br />