Chapter 4 Calibration Procedures<br />

34907A Analog Output<br />

34907A Analog Output<br />

Analog Output Verification Test<br />

This procedure is used to check the calibration of the analog outputs on<br />

the 34907A Multifunction Module. Install the module in slot 200.<br />

Verification checks are performed only for those output values with<br />

unique calibration constants.<br />

1 Make connections to analog output channels as shown below.<br />

To DMM<br />

34907A<br />

To<br />

DMM<br />

2 For each analog output, set each output value in the table below.<br />

Compare measurement results to the appropriate test limits shown in<br />

the table.<br />

DAC<br />

Output<br />

Quick<br />

Check<br />

0.000 V<br />

10.000 V Q<br />

Measured<br />

Output<br />

0.000 V<br />

10.000 V<br />

Error from Nominal<br />

1 year<br />

± 20 mV<br />

± 45 mV<br />

124