Chapter 3 Front-Panel Overview<br />
To Perform a Zero Adjustment<br />
To Perform a Zero Adjustment<br />
The instrument features closed case electronic calibration. No internal<br />
mechanical adjustments are required. The instrument calculates<br />
correction factors based upon an input reference value and stores the<br />
correction factors in non-volatile memory. This procedure demonstrates<br />
making the zero adjustment from the front panel. The gain adjustments<br />
are similar.<br />
DO NOT perform this procedure before reading Chapter 4. Chapter 4<br />
describes this procedure, the required input connections, input signals,<br />
and test considerations required for a valid adjustment.<br />
1 Configure the channel.<br />
You must configure a channel before applying performing the<br />
adjustment procedure. Configure the channel to DC VOLTS and 6½<br />
digits.<br />
2 Apply the input signal<br />
In this example, the input signal is a copper short (see page 76).<br />
3 Setup the calibration.<br />
The display will show PERFORM CAL.<br />
4 Set the adjustment value.<br />
The display will show the a number. Edit the number to the actual input<br />
value. For the Zero Adjustment, the input value is 0.000000.<br />
+000.000,000 mVDC<br />
4<br />
3<br />
5 Begin the adjustment.<br />
The display will show the progress of the adjustment. When all the<br />
adjustments are completed, the display will show done.<br />
DONE<br />
63