Chapter 3 Front-Panel Overview<br />

To Perform a Zero Adjustment<br />

To Perform a Zero Adjustment<br />

The instrument features closed case electronic calibration. No internal<br />

mechanical adjustments are required. The instrument calculates<br />

correction factors based upon an input reference value and stores the<br />

correction factors in non-volatile memory. This procedure demonstrates<br />

making the zero adjustment from the front panel. The gain adjustments<br />

are similar.<br />

DO NOT perform this procedure before reading Chapter 4. Chapter 4<br />

describes this procedure, the required input connections, input signals,<br />

and test considerations required for a valid adjustment.<br />

1 Configure the channel.<br />

You must configure a channel before applying performing the<br />

adjustment procedure. Configure the channel to DC VOLTS and 6½<br />

digits.<br />

2 Apply the input signal<br />

In this example, the input signal is a copper short (see page 76).<br />

3 Setup the calibration.<br />

The display will show PERFORM CAL.<br />

4 Set the adjustment value.<br />

The display will show the a number. Edit the number to the actual input<br />

value. For the Zero Adjustment, the input value is 0.000000.<br />

+000.000,000 mVDC<br />

4<br />

3<br />

5 Begin the adjustment.<br />

The display will show the progress of the adjustment. When all the<br />

adjustments are completed, the display will show done.<br />

DONE<br />

63