Chapter 4 Calibration Procedures<br />

Relay Verification<br />

Tests 4 - 43:<br />

Make the connections to the 34908A as shown in the diagram below.<br />

Connect copper shorts between all channels as shown. Be sure to route<br />

your wiring for proper strain relief and install the module cover. Install<br />

the 34908A in slot 200 of the <strong>34970A</strong>/<strong>34972A</strong>.<br />

Module Reference<br />

4<br />

Connections for 34908A Verification Tests 4 through 43<br />

For each test shown in the table starting on the next page, close only the<br />

channels shown in the “Channels Closed” column (closing one channel<br />

will open the previously closed channel). Record the 4-wire ohms<br />

measurements from the external DMM in the table.<br />

119