Chapter 4 Calibration Procedures<br />

Relay Verification<br />

...Continued from previous page<br />

Test<br />

#<br />

Channel<br />

Closed*<br />

External DMM Ohmmeter Connections<br />

HI LO HI<br />

Sense<br />

LO<br />

Sense<br />

Measured Value<br />

Test<br />

Limit<br />

Relay<br />

Measured<br />

18 Ch 8 P2 P2 P1 P1 ________Ohms 2.00 K408<br />

19 Ch 9 P2 P2 P1 P1 ________Ohms 2.00 K409<br />

Module Reference<br />

20 Ch 10 P2 P2 P1 P1 ________Ohms 2.00 K410<br />

21 Ch 11 P2 P2 P1 P1 ________Ohms 2.00 K411<br />

22 Ch 12 P2 P2 P1 P1 ________Ohms 2.00 K412<br />

23 Ch 13 P2 P2 P1 P1 ________Ohms 2.00 K413<br />

4<br />

24 Ch 14 P2 P2 P1 P1 ________Ohms 2.00 K414<br />

25 Ch 15 P2 P2 P1 P1 ________Ohms 2.00 K415<br />

26 Ch 16 P2 P2 P1 P1 ________Ohms 2.00 K416<br />

27 Ch 17 P2 P2 P1 P1 ________Ohms 2.00 K417<br />

28 Ch 18 P2 P2 P1 P1 ________Ohms 2.00 K418<br />

29 Ch 19 P2 P2 P1 P1 ________Ohms 2.00 K419<br />

30 Ch 21 P4 P4 P5 P5 ________Ohms — —<br />

31 Subtract (Test 30 - Test 4) ________Ohms 2.00 K522<br />

32 Ch 22 P4 P4 P5 P5 ________Ohms — —<br />

33 Subtract (Test 32 - Test 5) ________Ohms 2.00 K522<br />

*Only the channel currently under test should be closed at one time. All other channels should be open.<br />