Chapter 4 Calibration Procedures<br />

Input Connections<br />

Input Connections<br />

You will need an input multiplexer module to verify or adjust the<br />

internal DMM. Input connections can be made using a 34901A<br />

20-Channel Multiplexer.<br />

To use a 34901A to completely verify and adjust the internal DMM,<br />

make the following connections:<br />

Connections for<br />

DC/AC Current<br />

Connections for<br />

4-wire Ohms<br />

Connections for DC V, AC V,<br />

2- and 4-wire Ohms<br />

Copper Short<br />

34901A<br />

To<br />

Calibrator<br />

Note: Use shielded twisted pair PTFE insulated cables to reduce<br />

settling and noise errors. Connect the shield to the source LO output.<br />

You can also use a 34902A for test and adjustment of voltage, frequency,<br />

and resistance functions. You cannot test or adjust current inputs with a<br />

34902A. If you use a 34902A, connect the copper shorts to Channels 7<br />

and 15 and make the input connections to Channels 8 and 16.<br />