Chapter 4 Calibration Procedures<br />

Relay Verification<br />

Tests 2 - 3:<br />

Make the connections to the 34908A as shown in the diagram below. Be<br />

sure to route your wiring for proper strain relief and install the module<br />

cover. Install the 34908A in slot 200 of the <strong>34970A</strong>/<strong>34972A</strong>. Open all<br />

channels on the module by performing a Factory Reset (press and<br />

select “Recall State”; press again and select “Factory Reset”).<br />

Configure Channel 01 as follows: DC volts, 10 volt range, and 5½ digits.<br />

Connections for 34908A Verification Tests 2 and 3<br />

Enable reading monitoring by pressing on the selected channel (or<br />

use the ROUTe:MON command). Record the 4-wire ohms measurements<br />

from the external DMM in the following table.<br />

External DMM Ohmmeter Connections<br />

Test<br />

#<br />

Channel<br />

Configured<br />

HI LO HI<br />

Sense<br />

LO<br />

Sense<br />

Measured Value<br />

Test<br />

Limit<br />

Relay<br />

Measured<br />

2 Ch 01 H Com L Com H Com L Com ________Ohms — —<br />

3 Subtract (Test 2 - Test 1) ________Ohms 2.00 K421<br />

118