Chapter 4 Calibration Procedures<br />

Relay Verification<br />

Tests 9 - 10:<br />

Test<br />

#<br />

Channel<br />

Configured<br />

Open all channels on the module by performing a Factory Reset.<br />

Configure Channel 08 (module in slot 200) as follows: 4-wire ohms,<br />

1 k range, and 5½ digits.<br />

Enable reading monitoring by pressing on the selected channel (or<br />

use the ROUTe:MON command). Record the 4-wire ohms measurements<br />

from the external DMM in the following table.<br />

External DMM Ohmmeter Connections<br />

HI LO HI<br />

Sense<br />

LO<br />

Sense<br />

Measured Value<br />

Test<br />

Limit<br />

Relay<br />

Measured<br />

Module Reference<br />

9 Ch 08 P3 P3 P2 P2 ________Ohms — —<br />

10 Subtract (Test 9 - Test 4 ________Ohms 2.00 K328<br />

4<br />

Tests 11 - 27:<br />

Open all channels on the module by performing a Factory Reset. For<br />

each test, close only the channels shown in the “Channels Closed”<br />

column below (module in slot 200). Turn the Monitor Mode “off” and<br />

select “Banks Joined” from the Advanced menu. Record the 4-wire ohms<br />

measurements from the external DMM in the following table.<br />

External DMM Ohmmeter Connections<br />

Test<br />

#<br />

Channel<br />

Closed*<br />

HI LO HI<br />

Sense<br />

LO<br />

Sense<br />

Measured Value<br />

Test<br />

Limit<br />

Relay<br />

Measured<br />

11 Ch 16 & 1 P3 P3 P1 P1 ________Ohms 2.00 K301<br />

12 Ch 16 & 2 P3 P3 P1 P1 ________Ohms 2.00 K302<br />

13 Ch 16 & 3 P3 P3 P1 P1 ________Ohms 2.00 K303<br />

14 Ch 16 & 4 P3 P3 P1 P1 ________Ohms 2.00 K304<br />

15 Ch 16 & 5 P3 P3 P1 P1 ________Ohms 2.00 K305<br />

*Only the channel currently under test should be closed at one time. All other channels should be open.<br />