Chapter 6 Service<br />
Operating Checklist<br />
Operating Checklist<br />
Before returning your instrument to <strong>Agilent</strong> Technologies for service or<br />
repair, check the following items:<br />
Is the instrument inoperative?<br />
• Verify that the AC power cord is connected to the instrument.<br />
• Verify that the front-panel On/Standby switch has been pushed.<br />
4<br />
• Verify that the power-line fuse is installed and not open (see page 50).<br />
The instrument is shipped from the factory with a 500 mAT, 250 V<br />
fuse. This is the correct fuse for all line voltages.<br />
• Verify the power-line voltage setting. See “If the Instrument Does Not<br />
Turn On” on page 50.<br />
Does the instrument fail self-test?<br />
• Verify that the correct power-line voltage is selected. See “If the<br />
Instrument Does Not Turn On” on page 50.<br />
• Remove all input connections to the instrument. Errors may be<br />
induced by AC signals present on the input wiring during a self-test.<br />
Long test leads can act as an antenna causing pick-up of AC signals.<br />
Is the Current measurement function inoperative?<br />
• Verify the input protection fuses on the 34901A Multiplexer Module.<br />
If necessary, replace with a 1.5 A, 250 V NTD fuse.<br />
Note: The Current input is only available on channels 21 and 22 of the<br />
34901A module.<br />