Chapter 6 Service<br />

Operating Checklist<br />

Operating Checklist<br />

Before returning your instrument to <strong>Agilent</strong> Technologies for service or<br />

repair, check the following items:<br />

Is the instrument inoperative?<br />

• Verify that the AC power cord is connected to the instrument.<br />

• Verify that the front-panel On/Standby switch has been pushed.<br />

4<br />

• Verify that the power-line fuse is installed and not open (see page 50).<br />

The instrument is shipped from the factory with a 500 mAT, 250 V<br />

fuse. This is the correct fuse for all line voltages.<br />

• Verify the power-line voltage setting. See “If the Instrument Does Not<br />

Turn On” on page 50.<br />

Does the instrument fail self-test?<br />

• Verify that the correct power-line voltage is selected. See “If the<br />

Instrument Does Not Turn On” on page 50.<br />

• Remove all input connections to the instrument. Errors may be<br />

induced by AC signals present on the input wiring during a self-test.<br />

Long test leads can act as an antenna causing pick-up of AC signals.<br />

Is the Current measurement function inoperative?<br />

• Verify the input protection fuses on the 34901A Multiplexer Module.<br />

If necessary, replace with a 1.5 A, 250 V NTD fuse.<br />

Note: The Current input is only available on channels 21 and 22 of the<br />

34901A module.<br />