Chapter 3 Front-Panel Overview<br />

To Read the Totalizer Count<br />

To Read the Totalizer Count<br />

The multifunction module (34907A) has a 26-bit totalizer which can<br />

count pulses at a 100 kHz rate. You can manually read the totalizer<br />

count or you can configure a scan to read the count.<br />

1 Select the totalizer channel.<br />

Select the slot containing the multifunction module and continue turning<br />

the knob until TOTALIZE is displayed (channel 03).<br />

2 Configure the totalize mode.<br />

The internal count starts as soon as you turn on the instrument. You can<br />

configure the totalizer to reset the count to “0” after being read or it can<br />

count continuously and be manually reset.<br />

READ + RESET<br />

3 Read the count.<br />

The count is read once each time you press ; the count does not<br />

update automatically on the display. As configured in this example, the<br />

count is automatically reset to “0” each time you read it.<br />

12345 TOT<br />

The count will be displayed until you press another key, turn the knob,<br />

or until the display times out. To manually reset the totalizer count,<br />

press .<br />

Note: To add a totalizer channel to a scan list, press<br />

TOT READ choice.<br />

and select the<br />

68